HUNTLEY
Report
- Report Number
- MW1019106
- Event Type
- Injury
- Date Received
- June 9, 2000
- Date of Event
- February 29, 2000
- Report Date
- June 9, 2000
- Manufacturer
- HUNTLEIGH HEALTHCARE, INC.
- Product Code
- JOW
- Adverse Event
- Yes
- Report Source
- Voluntary report
- Reporter Location
- OR, US
- Reporter Occupation
- NURSE
Narratives
WHILE IN THE HOSP FOR ABDOMINAL SURGERY PT WORE ALP/SCD COMPRESSION BOOTS TO PREVENT DVT. WHEN THE BOOTS WERE REMOVED AFTER 2 DAYS PT HAD NUMBNESS IN CALF AND FOOT OF LEFT LEG. PT IS NOW BEING TREATED WITH PHYSICAL THERAPY TO TRY TO REGAIN FEELING IN THE LEG.
ADD'L INFO REC'D FROM MFR 7/24/00: MEDICAL CTR WAS CONTACTED REGARDING THE COMPLAINT FILED BY THE PT. A REP OF THE HOSP WAS AWARE OF THE COMPLAINTS MADE BY THE PT AND STATED THE HOSP DID NOT FEEL THE INJURY WAS DUE TO THE HUNTLEIGH FLOWTRON DVT PROPHYLAXIS INTERMITTENT COMPRESSION SYSTEM, AND THEREFORE DID NOT SEND THE PRODUCT BACK TO THE MFR AND HAD THE PRODUCT TESTED BY THEIR OWN SERVICES DEPT. NO MALFUNCTIONS WERE FOUND. THE FLOWTRON DVT PROPHYLAXIS SYSTEM WORKS BY INTERMITTENTLY COMPRESSING THE MUSCLES OF THE LEG, WHICH MIMICS THE ACTION OF WALKING, THEREBY HELPING TO PREVENT THROMBUS FORMATION. THE PUMP OPERATES ON A 60 SECOND CYCLE CONSISTING OF 12 SECONDS OF INFLATION FOLLOWED BY 48 SECONDS OF DEFLATION, ONE LEG AT A TIME. IT IS CO'S RECOMMENDATION THAT THE GARMENTS BE APPLIED SNUGLY TO THE PT'S EXTREMITIES TO ENSURE THE LEG IS FULLY COMPRESSED DURING INFLATION. IF THERE IS A PROBLEM WITH THE PUMP OR THE GARMENT, THE PUMPS AUDIBLE AND VISUAL ALARM WILL GO OFF. BASED UPON THE ABOVE, MFR DOES NOT BELIEVE THAT THE FLOWTRON SYSTEM CAUSED ANY INJURY TO THE PT. A FOLLOW-UP LETTER HAS BEEN SENT TO THE COMPLAINANT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HUNTLEY | ALP/SCD COMPRESSION BOOTS | JOW | HUNTLEIGH HEALTHCARE, INC. | * | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 41 YR | Other| R |