ATW MARKER GUIDEWIRE
Report
- Report Number
- 1016427-2012-00138
- Event Type
- Malfunction
- Date Received
- November 2, 2012
- Date of Event
- October 9, 2012
- Report Date
- October 9, 2012
- Manufacturer
- CORDIS CORPORATION
- Product Code
- DQX
- PMA / PMN Number
- K994358
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
THE REPORT RECEIVED FROM THE AFFILIATE INDICATED THAT AFTER THE SGW ATW .014 STR FLOPPY 195CM GUIDEWIRE WAS REMOVED FROM THE PACKAGING, THE DISTAL TIP WAS NOTED TO BE STRETCHED. THERE WAS NO REPORTED PATIENT INJURY. THE PHYSICIAN USED ANOTHER ONE TO COMPETE THE PROCEDURE SUCCESSFULLY. THE PRODUCT WAS STORED PROPERLY ACCORDING TO THE INSTRUCTIONS FOR USE/IFU. THERE WAS NO DAMAGE NOTED TO THE PRODUCT PACKAGING UPON INSPECTION PRIOR TO OPENING. THERE WAS NO REPORTED DIFFICULTY REMOVING THE PRODUCT FROM THE PACKAGING. THE PRODUCT PACKAGING WAS INSPECTED PRIOR TO OPENING WITH NO DAMAGE NOTED. THE PRODUCT WAS INSPECTED PRIOR TO USE AND APPEARED TO BE NORMAL. THE PRODUCT WAS PREPPED PROPERLY ACCORDING TO THE INSTRUCTIONS FOR USE (IFU) WITH NO PROBLEMS NOTED. THE FLEXIBLE, 'DELICATE' NATURE OF THE 'FLOPPY' TIP CONFIGURATION REQUIRES DUE CARE IN HANDLING AND USE, AS DIRECTED IN THE DEVICE INSTRUCTIONS FOR USE (IFU). ONE NON STERILE UNIT OF SGW ATW .014 STR FLOPPY 195CM WAS RECEIVED FOR ANALYSIS. THE DISTAL TIP PRESENTED AN UNRAVELING CONDITION. NO OTHER VISUAL DAMAGED WAS FOUND ON THE RECEIVED UNIT. SEE ATTACHED PICTURES. THE UNIT WAS OBSERVED UNDER MICROSCOPE AND NO OTHER ANOMALIES RATHER THAN THE VISUALLY DETECTED WERE OBSERVED. (B)(4). THE FAILURE REPORTED BY THE CUSTOMER AS 'DISTAL TIP - UNRAVELED/STRETCHED-PRIOR TO USE' WAS CONFIRMED AS THE RECEIVED CONDITION OF THE DEVICE, BUT IT IS NO POSSIBLE TO DETERMINE THE PLACE AND TIME OF THE DAMAGE, HOWEVER IT DOES NOT APPEARS TO BE MANUFACTURING RELATED OF THE PRODUCT, THE DEVICE DID NOT PRESENT ANY OBVIOUS INDICATIONS OF MANUFACTURING DEFECT OR ANOMALY THAT COULD BE CONTRIBUTED TO THE EVENTS AS REPORTED. THE RECORDS INDICATED THAT THE PRODUCT MET SPECIFICATION PRIOR TO SHIPMENT; THEREFORE NO CORRECTIVE OR PREVENTIVE ACTIONS WILL BE TAKEN AT THIS TIME.
THE PRODUCT WAS RETURNED FOR EVALUATION AND TESTING; HOWEVER, THE ENGINEERING EVALUATION IS NOT COMPLETE. ADDITIONAL INFORMATION WILL BE SUBMITTED WITHIN 30 DAYS UPON RECEIPT.
THE REPORT RECEIVED FROM THE AFFILIATE INDICATED THAT AFTER THE SGW ATW .014 STR FLOPPY 195 CM GUIDEWIRE WAS REMOVED FROM THE PACKAGING, THE DISTAL TIP WAS NOTED TO BE STRETCHED. THERE WAS NO REPORTED PATIENT INJURY. THE PHYSICIAN USED ANOTHER ONE TO COMPETE THE PROCEDURE SUCCESSFULLY. THE PRODUCT WAS STORED PROPERLY ACCORDING TO THE INSTRUCTIONS FOR USE/IFU. THERE WAS NO DAMAGE NOTED TO THE PRODUCT PACKAGING UPON INSPECTION PRIOR TO OPENING. THERE WAS NO REPORTED DIFFICULTY REMOVING THE PRODUCT FROM THE PACKAGING. THE PRODUCT PACKAGING WAS INSPECTED PRIOR TO OPENING WITH NO DAMAGE NOTED. THE PRODUCT WAS INSPECTED PRIOR TO USE AND APPEARED TO BE NORMAL. THE PRODUCT WAS PREPPED PROPERLY ACCORDING TO THE INSTRUCTIONS FOR USE (IFU) WITH NO PROBLEMS NOTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ATW MARKER GUIDEWIRE | CARDIOLOGY WIRES & METALS (DQX) | DQX | CORDIS CORPORATION | NA | F0511231 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |