ORTHO SUMMIT SAMPLE HANDLING SYSTEM
MDR report key: 281647
·
Received June 7, 2000
Report
- Report Number
- 2250051-2000-00164
- Event Type
- Malfunction
- Date Received
- June 7, 2000
- Date of Event
- April 20, 2000
- Report Date
- May 4, 2000
- Manufacturer
- HAMILTON BONADUZ AG
- Product Code
- JTC
- Product Problem
- Yes
- Report Source
- Distributor report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT WHILE RUNNING AN HEPATITIS C VIRUS ASSAY, SUMMIT SAMPLE HANDLER DID NOT PIPETTE THE CORRECT AMOUNT OF SAMPLE AND/OR DILUENT AND DID NOT GIVE AN ERROR MESSAGE. AN ORTHO FIELD SVC ENGINEER WAS DISPATCHED, AND COMPLAINT WAS CONFIRMED. FSE REPLACED TIPS, PLUNGER CLAMPS AND "LLD" SPRINGS. NO DEATH OR SERIOUS INJURY WAS ASSOCIATED WITH THIS EVENT. THIS REPORT CORRESPONDS TO ORTHO-CLINICAL DIAGNOSTICS COMPLAINT NUMBER 00-02207-05.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ORTHO SUMMIT SAMPLE HANDLING SYSTEM | SAMPLE HANDLER | JTC | HAMILTON BONADUZ AG | 135028 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Other |