PEAK PULSAR 2
Report
- Report Number
- 1226420-2012-00052
- Event Type
- Malfunction
- Date Received
- November 1, 2012
- Date of Event
- October 5, 2012
- Report Date
- January 22, 2013
- Manufacturer
- MEDTRONIC ADVANCED ENERGY, LLC
- Product Code
- MUL
- PMA / PMN Number
- K102029
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). EVAL, METHOD, RESULTS AND CONCLUSION: PRODUCT SCHEDULED FOR RETURN BUT NOT RECEIVED BY MANUFACTURER FOR INSPECTION. (B)(4).
PRODUCT EVENT: (B)(4). EVALUATION PROCESS: UNIT RECEIVED IN FAIR CONDITION AND INTERNAL VISUAL INSPECTION REVEALED NO LOOSE OR BROKEN COMPONENTS. UNIT DELIVERED RF ENERGY CORRECTLY INTO FIXED RESISTORS. ERROR LOG: 12 E3 (PATIENT RETURN ELECTRODE HAS POOR CONNECTION) THIS IS A NORMAL USE ERROR. TO ENSURE CUT AND COAG ENERGY DELIVERY WAS NOT OUT OF TOLERANCE/CALIBRATION STEPS 5.6 AND 5.12 OF FINAL TEST (TP-0050) WERE PERFORMED AND THE POWER MEASUREMENTS WERE WITHIN TOLERANCE AND THE COMPLAINT COULD NOT BE CONFIRMED. ROOT CAUSE: THE OUTPUT POWER WAS TESTED AND FOUND TO BE WITHIN SPECIFICATIONS. THE COMPLAINT OF OVER-POWERED OUTPUT ENERGY COULD NOT BE CONFIRMED. (B)(4).
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
SURGEONS STATE THAT ARCING IS OCCURRING WHILE DISSECTING IN COAG MODE AND THAT THE HAND PIECE SEEMS HOTTER THAN THE NORM CAUSING MORE TISSUE TO BE DAMAGED.
SURGEONS STATE THAT ARCING IS OCCURRING WHILE DISSECTING IN COAG MODE AND THAT THE HAND PIECE SEEMS HOTTER THAN THE NORM CAUSING MORE TISSUE TO BE DAMAGED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PEAK PULSAR 2 | GENERATOR,ELECTROSURGICAL,COAGULATION,CANCER | MUL | MEDTRONIC ADVANCED ENERGY, LLC | PS100-102 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |