FDA Adverse Event Malfunction Summary report: N

ACCESS

MDR report key: 2814872 · Received November 1, 2012

Report

Report Number
1416980-2012-03431
Event Type
Malfunction
Date Received
November 1, 2012
Date of Event
October 1, 2012
Report Date
October 5, 2012
Manufacturer
BAXTER HEALTHCARE
Product Code
FPA
PMA / PMN Number
K961225
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE IS NOT AVAILABLE FOR EVALUATION, THEREFORE, THE CONDITION CANNOT BE CONFIRMED NOR DUPLICATED AND THE ASSIGNABLE ROOT CAUSE COULD NOT BE DETERMINED. IF ADDITIONAL INFORMATION OR SAMPLES BECOME AVAILABLE, A FOLLOW UP REPORT WILL BE SUBMITTED. A BATCH REVIEW COULD NOT BE PERFORMED AS THE LOT NUMBER IS UNKNOWN.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

A CUSTOMER REPORTED TO BAXTER CORPORATE PRODUCT SURVEILLANCE A CLEARLINK NON-DEHP SECONDARY MEDICATION SET IN WHICH A BACKFLOW OCCURRED. ACCORDING TO THE REPORT, THE SECONDARY SET WAS ATTACHED TO THE UPPER MOST Y-SITE ON AN UNKNOWN CAREFUSION PRIMARY SET. THE PRIMARY SET WAS CONNECTED TO AN ALARIS INFUSION PUMP, AND THE SETTINGS WERE VERIFIED. THE NURSE STATED THAT THE PRIMARY BAG WAS LOWERED USING THE BLUE HANGER, WHICH WAS FULLY EXTENDED, INCLUDED WITH THE SECONDARY SET. THE SECONDARY SET WAS ATTACHED TO A MINI-BAG PLUS SOLUTION BAG. THE REPORTER INDICATED THAT THE SOLUTION FROM THE PRIMARY BAG WAS FLOWING INTO THE SECONDARY SOLUTION BAG. THE SECONDARY BAG BECAME SO FULL, IT LOOKED AS THOUGH IT WAS ABOUT TO BURST. THE FACILITY INDICATED THAT THEY SWITCHED TO CAREFUSION SECONDARY TUBING, AND NO FURTHER ISSUES HAVE BEEN OBSERVED. THE ALLEGED DEFECT OCCURRED DURING PATIENT USE. THERE WERE NO REPORTS OF PATIENT INJURY, ADVERSE EVENTS, OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ACCESS SET, ADMINISTRATION, INTRAVASCULAR FPA BAXTER HEALTHCARE

Patients

Seq Age Sex Outcome Treatment
1 ALARIS INFUSION PUMP| MINI-BAG PLUS| CAREFUSION PRIMARY SET