FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 2813629 · Received October 31, 2012

Report

Report Number
2649622-2012-16786
Event Type
Injury
Date Received
October 31, 2012
Date of Event
February 7, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S017
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED AND ANALYSIS FOUND NO ANOMALIES. HOWEVER, THE PROXIMAL CONDUCTOR HAD BLOOD/BODY FLUID (NOT OBSTRUCTED). THE OUTER INSULATION WAS MELTED, BREACHED CUT AND HAD A COSMETIC DEPRESSION. THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM. VISUAL ANALYSIS NOTED APPARENT EXPLANT DAMAGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ATRIAL LEAD HAD HIGH IMPEDANCE AND OVERSENSING WHICH TRIGGERED A PATIENT ALERT. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4076

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R D394TRG IMPLANTABLE PACEMAKER/CARDIO/DEFIB