FDA Adverse Event
Malfunction
Summary report: N
CARELINK PROGRAM W/TELEMETRY C
MDR report key: 2813607
·
Received October 31, 2012
Report
- Report Number
- 2182208-2012-03917
- Event Type
- Malfunction
- Date Received
- October 31, 2012
- Date of Event
- August 27, 2012
- Manufacturer
- MEDTRONIC, INC.
- Product Code
- DXY
- PMA / PMN Number
- P890003/S080
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): ECG (ELECTROCARDIOGRAM) CONNECTOR IS LOOSE ON A PRINTED CIRCUIT BOARD ASSEMBLY. IT IS NOTED THAT THE REPORTED WIRELESS COMMUNICATION ISSUES WERE DUE TO THE "ALLOW WIRELESS" RADIAL DIAL NOT BEING SELECTED ON THE PROGRAMMER. (B)(4): NO TELEMETRY WITH THE RF (RADIO FREQUENCY) HEAD; RF HEAD CABLE FOUND OUT OF ELECTRICAL SPECIFICATION.
Description of Event or Problem · 1
IT WAS REPORTED THE PROGRAMMER HAD ISSUES WITH WIRELESS COMMUNICATION. THE PROGRAMMER WAS RETURNED TO THE MANUFACTURER, ANALYZED AND TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CARELINK PROGRAM W/TELEMETRY C | PROGRAMMER | DXY | MEDTRONIC, INC. | 2090W |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other | 2290 PACING SYSTEM ANALYZER |