FDA Adverse Event
Injury
Summary report: N
KAPPA 700 DR
MDR report key: 2813584
·
Received October 31, 2012
Report
- Report Number
- 2647346-2012-01778
- Event Type
- Injury
- Date Received
- October 31, 2012
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- DXY
- PMA / PMN Number
- P980035
- Removal / Correction Number
- Z-2120-2011
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE DEVICE WAS RETURNED, ANALYZED, AND THE EMA WIREBOND WAS BROKEN.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE COULD NOT BE INTERROGATED (NO TELEMETRY). THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | KAPPA 700 DR | IMPLANTABLE PULSE GENERATOR | DXY | MEDTRONIC MED REL, INC. | KDR701 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 55 YR | Hospitalization| R | 4024 IMPLANTABLE PACING LEAD| 5076 IMPLANTABLE PACING LEAD |