FDA Adverse Event
Injury
Summary report: N
ATTAIN OTW
MDR report key: 2813569
·
Received October 31, 2012
Report
- Report Number
- 2649622-2012-17031
- Event Type
- Injury
- Date Received
- October 31, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P010015/S003
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD SHOWED OUT OF RANGE IMPEDANCE. THE RV TIP TO RV COIL MEASURED HIGH IMPEDANCE. THE RV LEAD HAD A DISTAL COIL FRACTURE. THE RV LEAD WAS CAPPED. IT WAS ALSO REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD IMPEDANCE WAS OUT OF RANGE. THE LEAD INTEGRITY WAS CHECKED DURING THE RV LEAD REVISION. THE LV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ATTAIN OTW | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 4193 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR | Hospitalization| R | D314TRG IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5076 IMPLANTABLE PACING LEAD |