FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2813529 · Received October 31, 2012

Report

Report Number
2649622-2012-17262
Event Type
Injury
Date Received
October 31, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
Z-0475-2011
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

PRODUCT EVALUATION SUMMARY: THE LEAD SEGMENTS WAS RETURNED TO THE MANUFACTURER AND ANALYZED AND NO ANOMALIES WERE FOUND. THE LEAD DISTAL ELECTRODE WAS COVERED IN BLOOD. THE LEAD INSULATION OVERLAY TUBING MELTED AND HAD ENVIRONMENTAL STRESS CRACKING (ESC.) THE LEAD OUTER INSULATION HAD COSMETIC DEPRESSION AND THE LEAD HAD APPARENT EXPLANT DAMAGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS NOISE PRESENT ON THE RIGHT VENTRICULAR (RV) LEAD. THEREFORE THE RV LEAD WAS REMOVED AND REPLACED WITH A NEW LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947

Patients

Seq Age Sex Outcome Treatment
1 77 YR Hospitalization| R (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5076 IMPLANTABLE PACING LEAD| 4194 IMPLANTABLE PACING LEAD