FDA Adverse Event Injury Summary report: N

ATTAIN ABILITY

MDR report key: 2813525 · Received October 31, 2012

Report

Report Number
2649622-2012-16768
Event Type
Injury
Date Received
October 31, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
OJX
PMA / PMN Number
P080006
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD NOISE, HIGH IMPEDANCE AND WAS FRACTURED. IT WAS ALSO REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD HAD HIGH THRESHOLD DUE TO EXIT BLOCK. THE RV LEAD WAS EXPLANTED AND REPLACED AND THE LV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN ABILITY IMPLANTABLE PACING LEAD OJX MEDTRONIC PUERTO RICO, INC. 4196

Patients

Seq Age Sex Outcome Treatment
1 64 YR Hospitalization| R D314TRG IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5071 IMPLANTABLE PACING LEAD