CAPSUREFIX NOVUS MRI
Report
- Report Number
- 2649622-2012-17259
- Event Type
- Injury
- Date Received
- October 31, 2012
- Date of Event
- September 11, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- NVN
- PMA / PMN Number
- P090013
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4). THE DEVICE WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. L(B)(4). THE DEVICE WAS RETURNED, ANALYZED AND THERE WAS BLOOD/BODY FLUID ON ALL CONDUCTORS (NOT OBSTRUCTED) . IT WAS NOTED THAT THE OUTER INSULATION WAS BREACHED CUT, THE HELIX/LOBE WAS DISTORTED/BENT, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM AND THERE WAS APPARENT EXPLANT DAMAGE. (B)(4). THE DEVICE WAS RETURNED, ANALYZED AND THERE WAS BLOOD/BODY FLUID ON ALL CONDUCTORS (NOT OBSTRUCTED). IT WAS NOTED THAT THE OUTER INSULATION WAS BREACHED CUT, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM AND THERE WAS APPARENT EXPLANT DAMAGE.
IT WAS REPORTED THAT THE DEVICE AND LEADS WERE REMOVED DUE TO INFECTION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX NOVUS MRI | IMPLANTABLE PACING LEAD | NVN | MEDTRONIC PUERTO RICO, INC. | 5086MRI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 80 YR | Hospitalization| R |