FDA Adverse Event Injury Summary report: N

CAPSUREFIX

MDR report key: 2813511 · Received October 31, 2012

Report

Report Number
2649622-2012-17019
Event Type
Injury
Date Received
October 31, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON ALL CONDUCTORS (NOT OBSTRUCTED) AND THE OUTER INSULATION HAD A COSMETIC DEPRESSION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD INCREASED IMPEDANCE AND LOSS OF VENTRICULAR CAPTURE AT MAXIMUM OUTPUT. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS WERE NOTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4068

Patients

Seq Age Sex Outcome Treatment
1 65 YR Hospitalization| R (B)(4) IMPLANTABLE PULSE GENERATOR| 4558M IMPLANTABLE PACING LEAD