CONCERTO CRT-D DR
Report
- Report Number
- 6000144-2012-06266
- Event Type
- Injury
- Date Received
- October 31, 2012
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- NIK
- PMA / PMN Number
- P010031/S031
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4). THE DEVICE WAS RETURNED AND ANALYZED. THE DEVICE LASTED 73% OF ITS EXPECTED LONGEVITY. BOTH LOADED AND UNLOADED PACING CURRENT DRAIN LEVELS WERE NORMAL FOR THIS DEVICE OVER THE RANGE OF BATTERY VOLTAGE IT OPERATED UNDER DURING ITS SERVICE TIME. THERE IS NO EVIDENCE TO INDICATE A PROBLEM WITH THE BATTERY. WITHOUT KNOWING THE PROGRAMMING HISTORY OF THIS DEVICE THERE IS NO WAY TO DETERMINE WHY IT DID NOT MEET ITS EXPECTED LONGEVITY CALCULATION.
THE DEVICE WAS REPLACED DUE TO EOL (END OF LIFE). THE DEVICE WAS RETURNED TO THE MANUFACTURER, ANALYZED, AND TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONCERTO CRT-D DR | IMPLANTABLE PACEMAKER/CARDIO/DEFIB | NIK | MEDTRONIC MED REL, INC. | C154DWK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 71 YR | Hospitalization| R | 4193 IMPLANTABLE PACING LEAD| 5076 IMPLANTABLE PACING LEAD| 6945 IMPLANTABLE TACHY LEAD |