FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2813459 · Received October 31, 2012

Report

Report Number
2649622-2012-16756
Event Type
Injury
Date Received
October 31, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
Z-0475-2011
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS A PATIENT ALERT TRIGGERED DUE TO HIGH RIGHT VENTRICULAR LEAD IMPEDANCE. IT WAS ALSO REPORTED THAT THERE WAS "INAPPROPRIATE EPISODE DETECTION DUE TO SENSING OF NOISE". THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947

Patients

Seq Age Sex Outcome Treatment
1 73 YR Hospitalization| R C174AWK IMPLANTABLE PACEMAKER/CARDIO/DEFIB