FDA Adverse Event Injury Summary report: N

CONCERTO CRT-D DR

MDR report key: 2813419 · Received October 31, 2012

Report

Report Number
6000144-2012-06589
Event Type
Injury
Date Received
October 31, 2012
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
NIK
PMA / PMN Number
P010031/S031
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY (B)(4) - THE DEVICE WAS RETURNED AND ANALYZED AND NO ANOMALIES WERE FOUND.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT HAD POCKET DISCOMFORT. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONCERTO CRT-D DR IMPLANTABLE PACEMAKER/CARDIO/DEFIB NIK MEDTRONIC MED REL, INC. C154DWK

Patients

Seq Age Sex Outcome Treatment
1 68 YR Hospitalization| R 4058M IMPLANTABLE PACING LEAD| 4193 IMPLANTABLE PACING LEAD| 6947 IMPLANTABLE TACHY LEAD