FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2813391 · Received October 31, 2012

Report

Report Number
2649622-2012-17008
Event Type
Injury
Date Received
October 31, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
Z-0475-2011
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT HAD AN EPISODE IN THE VENTRICULAR TACHYCARDIA (VT) ZONE AND THERAPY WAS WITHHELD DUE TO THE DEVICE DISCRIMINATOR. IT WAS ALSO REPORTED THAT THE PHYSICIAN FELT IT WAS A VT. IT WAS FURTHER REPORTED THAT THERE WAS FAR FIELD R-WAVE (FFRW) PRIOR TO DETECTION. IT WAS LATER REPORTED THAT THE PATIENT HAD BEEN OFF THEIR ANTIARRHYTHMICS DUE TO LACK OF INCOME TO OBTAIN THE MEDICINES AND THAT ARRHYTHMIAS OCCURRED DURING AN ACUTE MYOCARDIAL INFRACTION (MI) EPISODE IN THE EMERGENCY ROOM (ER.) THEREFORE THE VT DETECTION RATE WAS LOWERED AND HIGH RATE TIMEOUT WAS ENABLED. THE RIGHT VENTRICULAR (RV) LEAD AND DEVICE REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947

Patients

Seq Age Sex Outcome Treatment
1 55 YR Required Intervention 5076 IMPLANTABLE PACING LEAD