SCREW-IN
Report
- Report Number
- 2649622-2012-16752
- Event Type
- Injury
- Date Received
- October 31, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- K902002
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4). THE FULL LEAD WAS RETURNED IN SEGMENTS, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE OUTER INSULATION WAS BREACHED CUT, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM, THERE WAS TISSUE ON THE HELIX AND THERE WAS APPARENT EXPLANT DAMAGE.
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD HAD ELEVATED CAPTURE THRESHOLDS OVER TIME. IT WAS ALSO REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD HAD HIGH THRESHOLDS. THE RA LEAD WAS EXPLANTED AND REPLACED, AND THE LV LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SCREW-IN | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 5071 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR | Hospitalization| R | (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 6949 IMPLANTABLE TACHY LEAD |