FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 2813341 · Received October 31, 2012

Report

Report Number
2649622-2012-16745
Event Type
Injury
Date Received
October 31, 2012
Date of Event
August 31, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S017
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON THE DISTAL CONDUCTOR (NOT OBSTRUCTED), THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM AND THERE WAS TISSUE ON THE HELIX.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT WENT TO THE HOSPITAL BECAUSE OF DIZZINESS. THE RIGHT VENTRICULAR LEAD WAS FOUND TO SHOW VERY SMALL SENSING AND LOSS OF PACING. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4076

Patients

Seq Age Sex Outcome Treatment
1 86 YR Hospitalization| R SEDR01 IMPLANTABLE PULSE GENERATOR