FDA Adverse Event Malfunction Summary report: N

EXTERNAL PULSE GENERATOR

MDR report key: 2813286 · Received October 31, 2012

Report

Report Number
2183613-2012-01935
Event Type
Malfunction
Date Received
October 31, 2012
Manufacturer
MEDTRONIC MILACA INC.
Product Code
DTE
PMA / PMN Number
P820003/S72
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): ANALYSIS CONFIRMED THE REPORTED EVENT, THE RING COVER WAS BROKEN AND THE RING WAS MISSING. IT WAS ALSO NOTED THAT THE UPPER AND LOWER CASES WERE BROKEN AND CONTAMINATED, THE BATTERY RELEASE, LEAD FLEX COVER AND BATTERY DRAWER WERE CONTAMINATED, THE KEYBOARD PAD WAS COSMETICALLY DAMAGED, THE BATTERY CONTACTS WERE COMPRESSED AND THE TWO SIDE BAIL COVERS WERE CONTAMINATED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PACEMAKER HANDLE WAS BROKEN. THE DEVICE WAS RETURNED TO THE MANUFACTURER, ANALYZED, AND TESTED OUT OF SPECIFICATION. NO PATIENT INVOLVEMENT OR COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EXTERNAL PULSE GENERATOR DTE MEDTRONIC MILACA INC. 5388

Patients

Seq Age Sex Outcome Treatment
1 Other