PULSE-GENERATOR, PACEMAKER, EXTERNAL
Report
- Report Number
- 2183613-2012-01858
- Event Type
- Injury
- Date Received
- October 31, 2012
- Date of Event
- September 27, 2012
- Report Date
- November 9, 2012
- Manufacturer
- MEDTRONIC MILACA, INC.
- Product Code
- DTE
- PMA / PMN Number
- P820003/S70
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: ANALYSIS COULD NOT CONFIRM THE REPORTED EVENT, THE DEVICE PASSED FUNCTIONAL AND BENCH TESTING. IT WAS NOTED THAT THE UPPER AND LOWER CASES WERE BROKEN.
IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR INTERMITTENTLY SHUT DOWN AFTER A NEW BATTERY WAS USED. THE GENERATOR WAS RETURNED FOR SERVICE. IT WAS INDICATED ON THE RETURN PAPERWORK THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THIS EVENT.
IT WAS REPORTED THERE WAS INTERMITTENT SHUTDOWN AFTER NEW BATTERY WAS USED. THE DEVICE IS BEING RETURNED FOR REPAIR. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PULSE-GENERATOR, PACEMAKER, EXTERNAL | DTE | MEDTRONIC MILACA, INC. | 5348 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |