FDA Adverse Event Injury Summary report: N

CAPSUREFIX

MDR report key: 2813263 · Received October 31, 2012

Report

Report Number
2649622-2012-16509
Event Type
Injury
Date Received
October 31, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S2
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS A LEAD WARNING ON THE REMOTE MONITORING TRANSMISSION DUE TO LOW IMPEDANCE ON THE RIGHT ATRIAL (RA) LEAD. IT WAS ALSO REPORTED THAT THERE WAS NO P-WAVE SENSING NOTED VIA THE ANALYZER. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4568

Patients

Seq Age Sex Outcome Treatment
1 66 YR Hospitalization| R KDR701 IMPLANTABLE PULSE GENERATOR| 5076 IMPLANTABLE PACING LEAD