FDA Adverse Event Injury Summary report: N

VIRTUOSO DR

MDR report key: 2813213 · Received October 31, 2012

Report

Report Number
6000144-2012-06380
Event Type
Injury
Date Received
October 31, 2012
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
LWS
PMA / PMN Number
P980016/S062
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: A HIGH CURRENT DRAIN CONDITION WAS FOUND DURING DEVICE ANALYSIS. ELECTRICAL ANALYSIS FOUND THE CAUSE OF THE HIGH CURRENT DRAIN TO BE CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4). THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS ANALYZED AND REVEALED BATTERY VOLTAGE - PRE/APPROACHING ERI. THE DAILY BATTERY VOLTAGE TREND DATA IN SAVE TO DISK FILE (B)(4) SHOWS BATTERY = 2.89 TO 2.64 VOLTS BETWEEN (B)(6) 2012 AND (B)(6) 2012 AND IS BEFORE DEVICE RECOMMENDED REPLACEMENT TIME (RRT) <= 2.62 VOLT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WENT TO ELECTIVE REPLACEMENT INDICATOR EARLY AND THE DEVICE ONLY LASTED 4.4 YEARS. PREMATURE BATTERY DEPLETION IS SUSPECTED. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VIRTUOSO DR DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MEDTRONIC MED REL MEDTRONIC PUERTO RICO D154AWG

Patients

Seq Age Sex Outcome Treatment
1 80 YR Hospitalization| O| R 5076 IMPLANTABLE PACING LEAD| 6947 IMPLANTABLE TACHY LEAD