Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
OJX
PMA / PMN Number
P080006
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. IT WAS ALSO NOTED THAT ALL CONDUCTORS HAD BLOOD/BODY FLUID (NOT OBSTRUCTED), THE OUTER INSULATION WAS BREACHED CUT, AND THERE WAS APPARENT EXPLANT DAMAGE. (B)(4) THE DEVICE WAS RETURNED, ANALYZED, AND ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD HAD HIGH THRESHOLDS FOR THE FIRST SEVERAL MONTHS AND WAS EVENTUALLY PROGRAMMED OFF. THE DEVICE TRIGGERED ELECTIVE REPLACEMENT INDICATIOR EARLIER THEN EXPECTED. THE DEVICE AND LEAD WERE EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN ABILITY IMPLANTABLE PACING LEAD OJX MEDTRONIC PUERTO RICO, INC. 4196

Patients

Seq Age Sex Outcome Treatment
1 88 YR Hospitalization| R 6947 IMPLANTABLE TACHY LEAD| 1882TC COMPETITOR IMPLANTABLE PACING LEAD