FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2813143 · Received October 31, 2012

Report

Report Number
2649622-2012-16978
Event Type
Malfunction
Date Received
October 31, 2012
Date of Event
September 4, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
NVY
PMA / PMN Number
P920015/S055
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY FOR: THE FULL LEAD WAS RETURNED, ANALYZED AND THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANT ATTEMPT, THE PHYSICIAN NOTED ABNORMAL LOW BIPOLAR IMPEDANCE. IT WAS ALSO REPORTED THAT THERE MAY HAVE BEEN AN INSULATION BREACH. THE LEAD WAS WITHDRAWN FROM THE BODY AND A DIFFERENT LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD NVY MEDTRONIC PUERTO RICO, INC. 6947M

Patients

Seq Age Sex Outcome Treatment
1 71 YR Other (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 6947M IMPLANTABLE TACHY LEAD