FDA Adverse Event Injury Summary report: N

SPRINT FIDELIS

MDR report key: 2813099 · Received October 31, 2012

Report

Report Number
2649622-2012-16974
Event Type
Injury
Date Received
October 31, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S029
Removal / Correction Number
Z-0070-2008
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR [RV] LEAD WAS SENSING ATRIAL PACING (P WAVE) SPIKES AS R WAVES AND THESE WERE BEING DOUBLE COUNTED AS VENTRICULAR EVENTS. IT WAS ALSO REPORTED THAT THE RV LEAD HAD A HIGH NUMBER OF SHORT V-V INTERVALS, AN IMPEDANCE SPIKE, NOISE WAS PRESENT AND A LEAD FRACTURE WAS SUSPECTED. THE DETECTIONS WERE PROGRAMMED OFF AND PACING FUNCTIONS LEFT ON; THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT FIDELIS IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6949

Patients

Seq Age Sex Outcome Treatment
1 73 YR Required Intervention D153ATG IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5076 IMPLANTABLE PACING LEAD