FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE S

MDR report key: 2813069 · Received October 31, 2012

Report

Report Number
2649622-2012-16240
Event Type
Malfunction
Date Received
October 31, 2012
Date of Event
September 13, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S039
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM AND THERE WAS TISSUE ON THE HELIX. THE STYLET WAS BENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT, THE RIGHT VENTRICULAR LEAD (RV) HAD ATYPICAL IMPEDANCE MEASUREMENTS. A FRACTURE OF THE LEAD WAS SUSPECTED. THE LEAD WAS NOT USED AND A NEW RV LEAD WAS IMPLANTED DURING THE PROCEDURE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE S IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6935

Patients

Seq Age Sex Outcome Treatment
1 73 YR Other