FDA Adverse Event
Malfunction
Summary report: N
CARELINK
MDR report key: 2813043
·
Received October 31, 2012
Report
- Report Number
- 2182208-2012-03864
- Event Type
- Malfunction
- Date Received
- October 31, 2012
- Manufacturer
- MEDTRONIC, INC.
- Product Code
- KRG
- PMA / PMN Number
- P890003/S065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): RF (RADIO FREQUENCY) HEAD TELEMETRY BECOMES INTERMITTENT WHEN MOVING THE CABLE BACK AND FORTH AT THE POINT WHERE IT ENTERS THE BODY OF THE RF HEAD; RF HEAD CABLE FOUND OUT OF ELECTRICAL SPECIFICATION. RUBBER PORTION OF THE RF HEAD LABEL IS MISSING. LENS OF THE RF HEAD IS CRACKED.
Description of Event or Problem · 1
THE RF (RADIO FREQUENCY) HEAD WAS RETURNED TO THE MANUFACTURER, ANALYZED AND TESTED OUT OF SPECIFICATION. NO PATIENT INVOLVEMENT OR COMPLICATIONS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CARELINK | PROGRAMMER RF HEAD | KRG | MEDTRONIC, INC. | 2067 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |