CAPSUREFIX NOVUS
Report
- Report Number
- 2649622-2012-16720
- Event Type
- Injury
- Date Received
- October 31, 2012
- Report Date
- February 13, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P930039/S009
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
(B)(4).
PRODUCT EVENT SUMMARY: THE ACTUAL LEAD WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ALERTS: LEAD ALERT FOR RV Z > 3000 OHMS PRESENTED ON PROGRAMMER. IMPEDANCE/HIGH IMPEDANCE: RV LEAD IMPEDANCE 300-400 OHMS PRIOR TO (B)(4)-2012. STEADILY INCREASING OVER NEXT 34 WEEKS, BREAKING 3000 OHMS WEEK ENDING (B)(4)-2012. DAILY IMPEDANCES > 3K OHMS SINCE (B)(4)-2012.
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED A SLOW, STEADY RISE IN IMPEDANCE TO HIGH LEVELS. THE PATIENT WILL CONTINUE TO BE MONITORED, AND THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED A SLOW, STEADY RISE IN IMPEDANCE TO HIGH LEVELS. THE PATIENT WILL CONTINUE TO BE MONITORED, AND THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
IT WAS FURTHER REPORTED THAT THE LEAD WAS POSSIBLY FRACTURED. THE LEAD WAS CAPPED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX NOVUS | ELECTRODE, PACEMAKER, PERMANENT | DTB | MEDTRONIC PUERTO RICO, INC. | 5076 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 90 YR | Hospitalization| O| R | 5076 IMPLANTABLE PACING LEAD |