FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 2813042 · Received October 31, 2012

Report

Report Number
2649622-2012-16720
Event Type
Injury
Date Received
October 31, 2012
Report Date
February 13, 2013
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE ACTUAL LEAD WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ALERTS: LEAD ALERT FOR RV Z > 3000 OHMS PRESENTED ON PROGRAMMER. IMPEDANCE/HIGH IMPEDANCE: RV LEAD IMPEDANCE 300-400 OHMS PRIOR TO (B)(4)-2012. STEADILY INCREASING OVER NEXT 34 WEEKS, BREAKING 3000 OHMS WEEK ENDING (B)(4)-2012. DAILY IMPEDANCES > 3K OHMS SINCE (B)(4)-2012.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED A SLOW, STEADY RISE IN IMPEDANCE TO HIGH LEVELS. THE PATIENT WILL CONTINUE TO BE MONITORED, AND THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED A SLOW, STEADY RISE IN IMPEDANCE TO HIGH LEVELS. THE PATIENT WILL CONTINUE TO BE MONITORED, AND THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS FURTHER REPORTED THAT THE LEAD WAS POSSIBLY FRACTURED. THE LEAD WAS CAPPED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MEDTRONIC PUERTO RICO, INC. 5076

Patients

Seq Age Sex Outcome Treatment
1 90 YR Hospitalization| O| R 5076 IMPLANTABLE PACING LEAD