FDA Adverse Event Injury Summary report: N

ATTAIN OTW

MDR report key: 2813009 · Received October 31, 2012

Report

Report Number
2649622-2012-15979
Event Type
Injury
Date Received
October 31, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P010015/S012
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): NO ANOMALIES FOUND, HOWEVER THE PROXIMAL CONDUCTOR WAS DISTORTED, ALL CONDUCTORS WERE STRETCHED, THERE WAS BLOOD/BODY FLUID ON ALL CONDUCTORS (NOT OBSTRUCTED), THE OUTER INSULATION WAS MELTED, THE OUTER INSULATION WAS PULLED APART DUE TO OVERSTRESS, THE OUTER INSULATION HAD COSMETIC ENVIRONMENTAL STRESS CRACKING, THE INNER INSULATION WAS TORN, THERE WAS APPARENT EXPLANT DAMAGE AND THE LEAD WAS STRETCHED; FULL LEAD RETURNED AND ANALYZED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS AS LEAD DISLODGEMENT. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN OTW IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4194

Patients

Seq Age Sex Outcome Treatment
1 60 YR Hospitalization| R 6947 IMPLANTABLE TACHY LEAD| 5076 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB