FDA Adverse Event Malfunction Summary report: N

ATTAIN OTW

MDR report key: 2812983 · Received October 31, 2012

Report

Report Number
2649622-2012-17205
Event Type
Malfunction
Date Received
October 31, 2012
Date of Event
October 12, 2012
Report Date
November 8, 2012
Manufacturer
MPRI
Product Code
LWP
PMA / PMN Number
P010015/S012
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVENT SUMMARY FOR (B)(4): THE FULL LEAD WAS RETURNED AND ANALYZED. NO ANOMALIES WERE FOUND. THERE WAS BLOOD (NOT OBSTRUCTED) ON THE DISTAL AND PROXIMAL CONDUCTOR OF THE LEAD. THERE WAS AN OUTER INSULATION BREACH/CUT. THE PROXIMAL CONDUCTOR OF THE LEAD WAS DISTORTED, KINKED/BUCKLED. VISUAL ANALYSIS REVEALED THERE WAS APPARENT EXPLANT DAMAGE.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD WAS FOUND TO BE DISLODGED A DAY AFTER IMPLANT. THE LEAD WAS REMOVED AND REPLACED WITH A NEW LEAD. DURING REPLACEMENT WITH THE NEW THE LEAD THE LEAD WAS UNABLE TO FIXATE TO THE VEIN DUE TO THE LOBES NOT DEPLOYING SUCCESSFULLY. THE LEAD WAS REMOVED AND REPLACED WITH A NEW LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD WAS FOUND TO BE DISLODGED A DAY AFTER IMPLANT. THE LEAD WAS REMOVED AND REPLACED WITH A NEW LEAD. DURING REPLACEMENT OF THE NEW LEAD THE PHYSICIAN WAS UNABLE TO FIXATE TO THE VEIN DUE TO THE LOBES NOT DEPLOYING SUCCESSFULLY. THE LEAD WAS REMOVED AND REPLACED WITH A NEW LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN OTW PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE LWP MPRI 4194

Patients

Seq Age Sex Outcome Treatment
1 59 YR Hospitalization| R (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 6935 IMPLANTABLE TACHY LEAD| 4076 IMPLANTABLE PACING LEAD