FDA Adverse Event
Malfunction
Summary report: N
CARELINK
MDR report key: 2812976
·
Received October 31, 2012
Report
- Report Number
- 2182208-2012-03857
- Event Type
- Malfunction
- Date Received
- October 31, 2012
- Manufacturer
- MEDTRONIC, INC.
- Product Code
- DXY
- PMA / PMN Number
- P890003/S065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) ANALYZER IS FUNCTIONALLY OK. PATIENT CABLE CONNECTION IS BROKEN. LABEL IS PULLING LOOSE.
Description of Event or Problem · 1
THE PROGRAMMER ANALYZER WAS RETURNED TO THE MANUFACTURER, ANALYZED AND TESTED OUT OF SPECIFICATION. NO PATIENT INVOLVEMENT OR COMPLICATIONS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CARELINK | PACING SYSTEM ANALYZER | DXY | MEDTRONIC, INC. | 2290 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |