SPRINT QUATTRO SECURE
Report
- Report Number
- 2649622-2012-15974
- Event Type
- Injury
- Date Received
- October 31, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S24
- Removal / Correction Number
- Z-0475-2011
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE PARTIAL LEAD WAS RETURNED IN SEGMENTS AND ANALYZED. NO ANOMALIES WERE FOUND. THE DEFIB CONDUCTOR WAS DISTORTED. ALL CONDUCTORS HAD BLOOD/BODY FLUID (NOT OBSTRUCTED). THE OUTER INSULATION HAD COSMETIC ENVIRONMENTAL STRESS CRACKING, WAS BREACHED CUT, AND HAD COSMETIC DEPRESSION. THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM. VISUAL ANALYSIS NOTED THE LEAD HAD APPARENT EXPLANT DAMAGE.
IT WAS REPORTED THAT THE RV (RIGHT VENTRICULAR) THRESHOLD WAS INCREASING STEADILY, AND AN X-RAY SHOWED AN APPARENT LEAD DISLODGEMENT. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6947 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | Hospitalization| R | D274VRC IMPLANTABLE PACEMAKER/CARDIO/DEFIB |