FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2812973 · Received October 31, 2012

Report

Report Number
2649622-2012-15974
Event Type
Injury
Date Received
October 31, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
Z-0475-2011
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE PARTIAL LEAD WAS RETURNED IN SEGMENTS AND ANALYZED. NO ANOMALIES WERE FOUND. THE DEFIB CONDUCTOR WAS DISTORTED. ALL CONDUCTORS HAD BLOOD/BODY FLUID (NOT OBSTRUCTED). THE OUTER INSULATION HAD COSMETIC ENVIRONMENTAL STRESS CRACKING, WAS BREACHED CUT, AND HAD COSMETIC DEPRESSION. THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM. VISUAL ANALYSIS NOTED THE LEAD HAD APPARENT EXPLANT DAMAGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RV (RIGHT VENTRICULAR) THRESHOLD WAS INCREASING STEADILY, AND AN X-RAY SHOWED AN APPARENT LEAD DISLODGEMENT. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947

Patients

Seq Age Sex Outcome Treatment
1 66 YR Hospitalization| R D274VRC IMPLANTABLE PACEMAKER/CARDIO/DEFIB