Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: (B)(4): THE DISTAL PORTION OF THE LEAD WAS RETURNED, ANALYZED AND THE OUTER INSULATION WAS BREACHED AND HAD A DEPRESSION. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON ALL CONDUCTORS (NOT OBSTRUCTED), THE OUTER INSULATION WAS BREACH CUT, THE HELIX/LOBE WAS DISTORTED/BENT AND THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM. THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED, THE PATIENT DEVELOPED AN INFECTION. THE LEAD WAS RETURNED, ANALYZED AND THE RIGHT ATRIAL LEAD SUBSEQUENTLY TESTED OUT OF SPECIFICATION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076

Patients

Seq Age Sex Outcome Treatment
1 69 YR Hospitalization| R 6947 IMPLANTABLE TACHY LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB