FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2812844 · Received October 31, 2012

Report

Report Number
2649622-2012-16216
Event Type
Injury
Date Received
October 31, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD SUPERIOR VENA CAVA (SVC) IMPEDANCE WAS ERRATIC AND HIGH, AND THAT THE HIGH VOLTAGE-B (HVB) IMPEDANCE ROSE. IT WAS ALSO REPORTED THAT THE RV LEAD THRESHOLD HAS BEEN CHRONICALLY ELEVATED SINCE IMPLANT WITH T-WAVE OVERSENSING (TWOS) IN THE PAST THAT WAS REPROGRAMMED AROUND. THE RV LEAD WAS EXPLANTED AND REPLACED WITH A NEW LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947

Patients

Seq Age Sex Outcome Treatment
1 74 YR Hospitalization| R D154VWC IMPLANTABLE PACEMAKER/CARDIO/DEFIB