CAPSUREFIX NOVUS

MDR report key: 2812828 · Received October 31, 2012

Report

Report Number
2649622-2012-16952
Event Type
Malfunction
Date Received
October 31, 2012
Date of Event
October 9, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S017
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED AND ANALYZED. NO ANOMALIES WERE FOUND. THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT, WHEN THE SCREW ON THE LEAD WAS DEPLOYED, THE LEAD WOULD TURN AS WELL. THE LEAD DISLODGED MULTIPLE TIMES WHILE ATTEMPTING TO DEPLOY THE SCREW. THE LEAD WAS REMOVED AND ANOTHER LEAD WAS IMPLANTED INSTEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4076

Patients

Seq Age Sex Outcome Treatment
1 86 YR Other