FDA Adverse Event Injury Summary report: N

VIRTUOSO VR

MDR report key: 2812826 · Received October 31, 2012

Report

Report Number
6000144-2012-06010
Event Type
Injury
Date Received
October 31, 2012
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
LWS
PMA / PMN Number
P980016/S062
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) ANALYSIS OF THE DEVICE FOUND A HIGH CURRENT DRAIN CONDITION. ELECTRICAL ANALYSIS REVEALED THE CAUSE OF THE HIGH CURRENT DRAIN WAS CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. TIME OF RRT IN SAVE TO DISK ON (B)(6) 2012 DEVICE RRT <=2.62 VOLT. TREND DATA SHOWS MIN BAT=2.64 TO 2.62 VOLTS BETWEEN (B)(6) 2012. TWO PATIENT ALERTS FOR LOW BATTERY VOLTAGE OCCURRED ON (B)(6) 2012 02:15:03 AND (B)(6) 2012 02:15:03.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WAS AT RECOMMENDED REPLACEMENT TIME (RRT) AND HAD EARLY BATTERY DEPLETION. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VIRTUOSO VR IMPLANTABLE PACEMAKER/CARDIO/DEFIB LWS MEDTRONIC MED REL, INC. D154VWC

Patients

Seq Age Sex Outcome Treatment
1 29 YR Hospitalization| R 6945 IMPLANTABLE TACHY LEAD