FDA Adverse Event Malfunction Summary report: N

EXTERNAL PULSE GENERATOR

MDR report key: 2812779 · Received October 31, 2012

Report

Report Number
2183613-2012-01902
Event Type
Malfunction
Date Received
October 31, 2012
Date of Event
October 15, 2012
Manufacturer
MEDTRONIC MILACA INC.
Product Code
DTE
PMA / PMN Number
P820003/S72
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED BY THE BIOMEDICAL ENGINEER THAT THE EXTERNAL PULSE GENERATOR (EPG) WOULD POWER ON, BUT THEN SHUT OFF. IT WAS FURTHER REPORTED THAT THE DEVICE POWERED ON FINE WITH A NEW BATTERY. TECHNICAL SUPPORT (TS) RECOMMENDED CHECKING OUT THE DEVICE BEFORE ISSUING BACK TO THE FLOOR. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EXTERNAL PULSE GENERATOR DTE MEDTRONIC MILACA INC. 5388

Patients

Seq Age Sex Outcome Treatment
1 Other