FDA Adverse Event
Malfunction
Summary report: N
CARELINK
MDR report key: 2812691
·
Received October 31, 2012
Report
- Report Number
- 2182208-2012-03848
- Event Type
- Malfunction
- Date Received
- October 31, 2012
- Date of Event
- October 18, 2012
- Manufacturer
- MEDTRONIC, INC.
- Product Code
- KRG
- PMA / PMN Number
- P890003/S065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PROGRAMMER RADIO FREQUENCY HEAD WAS NOT GETTING TELEMETRY, BUT ONLY WITH PACEMAKERS. THE HEAD WAS RETURNED FOR SERVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CARELINK | PROGRAMMER RF HEAD | KRG | MEDTRONIC, INC. | 2067L |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |