FDA Adverse Event Injury Summary report: N

CAPSUREFIX

MDR report key: 2812680 · Received October 31, 2012

Report

Report Number
2649622-2012-15710
Event Type
Injury
Date Received
October 31, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT WAS HAVING HEART PALPITATIONS AT NIGHT. A LEAD WARNING WAS OBSERVED AND UNIPOLAR IMPEDANCES WERE HIGHER THAN BIPOLAR IMPEDANCES. THE LEAD WAS REPROGRAMMED AND REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4068

Patients

Seq Age Sex Outcome Treatment
1 80 YR Required Intervention KDR403 IMPLANTABLE PULSE GENERATOR| 4068 IMPLANTABLE PACING LEAD