SPRINT QUATTRO SECURE S
Report
- Report Number
- 2649622-2012-16208
- Event Type
- Injury
- Date Received
- October 31, 2012
- Report Date
- January 8, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S039
- Removal / Correction Number
- Z-0474-2011
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GB
- Reporter Occupation
- PHYSICIAN
Narratives
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS.
PRODUCT EVENT SUMMARY - THE FULL LEAD WAS RETURNED IN SEGMENTS AND ANALYZED. THERE WERE NO ANOMALIES FOUND. IT WAS NOTED THAT THE OUTER INSULATION HAD A COSMETIC DEPRESSION ALONG WITH TUBING OVERLAY ENVIRONMENTAL STRESS CRACKING. THE DISTAL CONDUCTOR WAS COVERED WITH BLOOD (NOT OBSTRUCTED).
IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE THERAPY. THE RIGHT VENTRICULAR LEAD HAD OVERSENSING OF NON-PHYSIOLOGIC SIGNALS. THE LEAD DETECTIONS WERE PROGRAMMED OFF AND THE LEAD WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE S | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MEDTRONIC PUERTO RICO, INC. | 6935 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 71 YR | Hospitalization| L| R |