FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE S

MDR report key: 2812669 · Received October 31, 2012

Report

Report Number
2649622-2012-16208
Event Type
Injury
Date Received
October 31, 2012
Report Date
January 8, 2013
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S039
Removal / Correction Number
Z-0474-2011
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GB
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY - THE FULL LEAD WAS RETURNED IN SEGMENTS AND ANALYZED. THERE WERE NO ANOMALIES FOUND. IT WAS NOTED THAT THE OUTER INSULATION HAD A COSMETIC DEPRESSION ALONG WITH TUBING OVERLAY ENVIRONMENTAL STRESS CRACKING. THE DISTAL CONDUCTOR WAS COVERED WITH BLOOD (NOT OBSTRUCTED).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE THERAPY. THE RIGHT VENTRICULAR LEAD HAD OVERSENSING OF NON-PHYSIOLOGIC SIGNALS. THE LEAD DETECTIONS WERE PROGRAMMED OFF AND THE LEAD WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE S DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MEDTRONIC PUERTO RICO, INC. 6935

Patients

Seq Age Sex Outcome Treatment
1 71 YR Hospitalization| L| R