FDA Adverse Event Injury Summary report: N

ENRHYTHM DR

MDR report key: 2812641 · Received October 31, 2012

Report

Report Number
6000144-2012-06362
Event Type
Injury
Date Received
October 31, 2012
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
DXY
PMA / PMN Number
P980035/S38
Removal / Correction Number
Z-1440-2010
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WENT TO ELECTIVE REPLACEMENT INDICATOR EARLY AND PREMATURE BATTERY DEPLETION IS SUSPECTED. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENRHYTHM DR IMPLANTABLE PULSE GENERATOR DXY MEDTRONIC MED REL, INC. P1501DR

Patients

Seq Age Sex Outcome Treatment
1 62 YR Hospitalization| R 5076 X2 IMPLANTABLE PACING LEADS