CAPSUREFIX NOVUS MRI

MDR report key: 2812632 · Received October 31, 2012

Report

Report Number
2649622-2012-15950
Event Type
Injury
Date Received
October 31, 2012
Date of Event
January 26, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
NVN
PMA / PMN Number
P090013
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED AND ANALYSIS FOUND NO ANOMALIES. HOWEVER, IT WAS NOTED THAT THE PROXIMAL CONDUCTOR WAS DISTORTED AND HAD BLOOD/BODY FLUID (NOT OBSTRUCTED). THE INNER INSULATION WAS KINKED BUCKLED. THE INNER TUBING WAS KINKED BUCKLED. THE OUTER INSULATION HAD A COSMETIC CUT, WAS BREACHED CUT AND HAD A COSMETIC DEPRESSION. THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM. THERE WAS TISSUE ON THE HELIX AND THE LEAD WAS STRETCHED. VISUAL ANALYSIS NOTED THAT THERE WAS APPARENT EXPLANT DAMAGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD HAD DECREASED SENSING OVER TIME SINCE IMPLANT. AN X-RAY SHOWED THAT SOME OF THE LEAD SLACK HAD BEEN TAKEN UP. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS MRI IMPLANTABLE PACING LEAD NVN MEDTRONIC PUERTO RICO, INC. 5086MRI

Patients

Seq Age Sex Outcome Treatment
1 68 YR Hospitalization| R (B)(4) IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PULSE GENERATOR