FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO

MDR report key: 2812631 · Received October 31, 2012

Report

Report Number
2649622-2012-17445
Event Type
Injury
Date Received
October 31, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S17
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD TRIGGERED AN ALERT FOR HIGH IMPEDANCE. THERE WAS A SPIKE IN PACING IMPEDANCE; CAPTURE THRESHOLD HAD INCREASED, AND THE LEAD WAS FRACTURED. THE DETECTIONS WERE TURNED OFF AND THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6944

Patients

Seq Age Sex Outcome Treatment
1 89 YR Required Intervention (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5592 IMPLANTABLE PACING LEAD