FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 2812622 · Received October 31, 2012

Report

Report Number
2649622-2012-17444
Event Type
Malfunction
Date Received
October 31, 2012
Date of Event
September 27, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVALUATION SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS ALSO NOTED THAT THE DISTAL LOW VOLTAGE ELECTRODE HAD BLOOD AND THE DISTAL END APPEARED TO HAVE DAMAGE. ALSO THE LEAD APPEARED TO BE STRETCHED AND DAMAGED AT IMPLANT.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT, THE RIGHT VENTRICULAR (RV) LEAD HAD A CONSISTENTLY HIGH IMPEDANCE READING OF APPROXIMATELY 1500 TO 1700 OHMS AND POOR SENSING AT APPROXIMATELY 5 MILLIVOLTS. THE RV LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MEDTRONIC PUERTO RICO, INC. 5076

Patients

Seq Age Sex Outcome Treatment
1 75 YR Other