Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) - THE FULL LEAD WAS RETURNED AND ANALYZED. NO ANOMALIES WERE FOUND. BLOOD/BODY FLUID WAS FOUND ON THE DISTAL CONDUCTOR (NOT OBSTRUCTED), AND BLOOD WAS FOUND ON THE TIP OF THE ELECTRODE. THE OUTER INSULATION WAS MELTED, AND EXHIBITED A COSMETIC CUT, DEPRESSION, AND COSMETIC ENVIRONMENTAL STRESS CRACKING. THE LEAD EXHIBITED APPARENT EXPLANT DAMAGE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE LEFT VENTRICULAR [LV] LEAD HAD DISLODGED AND WAS NOT CAPTURING. THE LV LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.