FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS MRI

MDR report key: 2812424 · Received October 31, 2012

Report

Report Number
2649622-2012-15935
Event Type
Injury
Date Received
October 31, 2012
Date of Event
July 30, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
NVN
PMA / PMN Number
P090013
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE DEVICE WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. (B)(4)THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE PROXIMAL CONDUCTOR WAS DISTORTED, THERE WAS BLOOD/BODY FLUID ON THE PROXIMAL CONDUCTOR (NOT OBSTRUCTED), THE OUTER INSULATION HAD A COSMETIC CUT, THE OUTER INSULATION WAS BREACHED CUT, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM AND THERE WAS APPARENT EXPLANT DAMAGE. (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON THE PROXIMAL CONDUCTOR (NOT OBSTRUCTED), THE INNER INSULATION WAS KINKED/BUCKLED, THE INNER TUBING WAS KINKED/BUCKLED, THE OUTER INSULATION HAD A COSMETIC CUT, THE OUTER INSULATION WAS BREACHED CUT, THE OUTER INSULATION HAD A COSMETIC DEPRESSION, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM, THERE WAS TISSUE PRESENT ON HELIX, THERE WAS APPARENT EXPLANT DAMAGE AND THE LEAD WAS STRETCHED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE AND LEADS WERE REMOVED DUE TO INFECTION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS MRI IMPLANTABLE PACING LEAD NVN MEDTRONIC PUERTO RICO, INC. 5086MRI

Patients

Seq Age Sex Outcome Treatment
1 80 YR Hospitalization| R