FDA Adverse Event
Malfunction
Summary report: N
SPRINT QUATTRO SECURE S
MDR report key: 2812396
·
Received October 31, 2012
Report
- Report Number
- 2649622-2012-16924
- Event Type
- Malfunction
- Date Received
- October 31, 2012
- Date of Event
- October 4, 2012
- Report Date
- October 24, 2012
- Manufacturer
- MPRI
- Product Code
- LWS
- PMA / PMN Number
- P920015/S039
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS.
Additional Manufacturer Narrative · 1
PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED. THE HELIX WAS BENT. IT WAS VISUALLY NOTED THAT THE LEAD WAS DAMAGED AT IMPLANT.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING IMPLANT, THE HELIX SCREW MECHANISM DID NOT WORK PROPERLY. THE LEAD WAS NOT USED AND A DIFFERENT LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE S | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MPRI | 6935 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |