FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2812328 · Received October 31, 2012

Report

Report Number
2649622-2012-16435
Event Type
Malfunction
Date Received
October 31, 2012
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
Z-0475-2011
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) - THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. THE LEAD INTEGRITY ALERT TRIGGERED. THERE WAS ONE PATIENT ALERT FOR LEAD FAILURE PREDICTOR ON (B)(6) 2012. THERE WAS OVERSENSING WITH ELEVEN VENTRICULAR NON-SUSTAINED TACHYCARDIA EPISODES LESS THAN OR EQUAL TO 210 MILLISECONDS ON (B)(6) 2012. THE VENTRICULAR SHORT INTERVAL COUNT IS EQUAL TO 331 COUNTS, IN 0.17 DAY, ON (B)(6) 2012.

Description of Event or Problem · 1

THE ANALYSIS WAS REVIEWED. THE LEAD TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. THE LEAD IS STILL IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947

Patients

Seq Age Sex Outcome Treatment
1 59 YR Other (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 4194 IMPLANTABLE PACING LEAD| 5076 IMPLANTABLE PACING LEAD