SPRINT QUATTRO SECURE
Report
- Report Number
- 2649622-2012-16435
- Event Type
- Malfunction
- Date Received
- October 31, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S24
- Removal / Correction Number
- Z-0475-2011
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) - THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. THE LEAD INTEGRITY ALERT TRIGGERED. THERE WAS ONE PATIENT ALERT FOR LEAD FAILURE PREDICTOR ON (B)(6) 2012. THERE WAS OVERSENSING WITH ELEVEN VENTRICULAR NON-SUSTAINED TACHYCARDIA EPISODES LESS THAN OR EQUAL TO 210 MILLISECONDS ON (B)(6) 2012. THE VENTRICULAR SHORT INTERVAL COUNT IS EQUAL TO 331 COUNTS, IN 0.17 DAY, ON (B)(6) 2012.
THE ANALYSIS WAS REVIEWED. THE LEAD TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. THE LEAD IS STILL IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6947 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR | Other | (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 4194 IMPLANTABLE PACING LEAD| 5076 IMPLANTABLE PACING LEAD |