FDA Adverse Event
Malfunction
Summary report: N
CAPSUREFIX NOVUS MRI
MDR report key: 2812323
·
Received October 31, 2012
Report
- Report Number
- 2649622-2012-16180
- Event Type
- Malfunction
- Date Received
- October 31, 2012
- Date of Event
- September 14, 2012
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- NVN
- PMA / PMN Number
- P090013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4). THE FULL LEAD WAS RETURNED AND ANALYZED. NO ANOMALIES WERE FOUND. BLOOD WAS FOUND IN/ON THE HELIX/LOBE MECHANISM.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE THE LEAD HAD HIGH IMPEDANCE. MULTIPLE ATTEMPTS WERE MADE TO MEASURE THE LEAD, INCLUDING USING NEW ANALYZER CABLES, BUT THE IMPEDANCES REMAINED HIGH. THE LEAD WAS REMOVED AND REPLACED WITH A NEW LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX NOVUS MRI | IMPLANTABLE PACING LEAD | NVN | MEDTRONIC PUERTO RICO, INC. | 5086MRI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 79 YR | Other |